4.7 Article

Environmental and biological determination of acrolein using new cold fiber solid phase microextraction with gas chromatography mass spectrometry

Journal

ANALYTICAL AND BIOANALYTICAL CHEMISTRY
Volume 409, Issue 11, Pages 2821-2828

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00216-017-0226-9

Keywords

Acrolein; Cold fiber SPME; Indoor; Exhaled air

Funding

  1. Fundacao de Amparo a Pesquisa de Minas Gerais (FAPEMIG)
  2. Conselho Nacional de Desenvolvimento Cientifico e Tecnologico (CNPq)
  3. Coordenacao de Aperfeicoamento de Pessoal de Nivel Superior (Capes)

Ask authors/readers for more resources

Acrolein is a pollutant released daily to the indoor environment from different sources. The present study reports the development of a simple and sensitive cold fiber solid phase microextraction sampling method for the determination of acrolein in exhaled air and indoor air by gas chromatography mass spectrometry. O-(2,3,4,5,6-pentafluorobenzyl) hydroxylamine was used as derivatizing agent supported on a 65-mu m polydimethylsiloxane-divinylbenzene SPME fiber. An acrolein permeation tube at 326.25 ng min(-1) rate was used to generate gaseous standards. The method shows good results for main validation parameters. The limits of detection and quantification were 2.88 and 5.08 mu g m(-3), respectively, for indoor analysis; and 2.40 and 3.79 mu g m(-3), respectively, for exhaled air analysis. The precision showed standard deviation ranges from 6.00 to 8.00% for intra-assay analyses and from 8.00 to 10.00% for inter-assay analyses. After optimizing the conditions, analyses of real sampleswere performed on indoor environments contaminated by cigarette smoke, or heated oil, including pastry shops, restaurants, churros stands, and closed parking cars located in the city of Belo Horizonte, Brazil. Acrolein breaths of exposed people were also determined. A good Pearson correlation coefficient (r = 0.901) was observed between the concentration of acrolein in indoor air and exhaled air, allowing to propose acrolein breath as environmental exposure biomarker.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available