4.5 Article

Quantitative Assessment of Variation in CT Parameters on Texture Features: Pilot Study Using a Nonanatomic Phantom

Journal

AMERICAN JOURNAL OF NEURORADIOLOGY
Volume 38, Issue 5, Pages 981-985

Publisher

AMER SOC NEURORADIOLOGY
DOI: 10.3174/ajnr.A5139

Keywords

-

Funding

  1. KAKEN [26861033]
  2. Toshiba Medical Systems
  3. Bracco Diagnostics
  4. Grants-in-Aid for Scientific Research [26861033] Funding Source: KAKEN

Ask authors/readers for more resources

Our aim was to evaluate 'changes in texture features based on variations in CT parameters on a phantom. Scans were performed with varying milliampere, kilovolt, section thickness, pitch, and acquisition mode. Forty-two texture features were extracted by using an in-house-developed Matlab program. Two-tailed t tests and false-detection analyses were performed with significant differences in texture features based on detector array configurations (Q values = 0.001-0.006), section thickness (Q values = 0.0002-0.001), and acquisition mode (Q values = 0.003-0.006). Variations in milliampere and kilovolt had no significant effect.