4.7 Article

A prospective randomized evaluation of the TriGuard™ HDH embolic DEFLECTion device during transcatheter aortic valve implantation: results fromthe DEFLECT III trial

Journal

EUROPEAN HEART JOURNAL
Volume 36, Issue 31, Pages 2070-2078

Publisher

OXFORD UNIV PRESS
DOI: 10.1093/eurheartj/ehv191

Keywords

Neuroprotection; Transcatheter aortic valve implantation; Diffusion-weighted imaging; Cerebral ischaemia; Stroke prevention

Funding

  1. Keystone Heart, Ltd.
  2. National Institute of Health Research Bristol Cardiovascular Biomedical Research Unit

Ask authors/readers for more resources

Aims To evaluate the safety, efficacy, and performance of the TriGuard (TM) HDH Embolic Deflection Device (TriGuard) compared with no cerebral protection in patients undergoing transcatheter aortic valve implantation (TAVI). Methods and results From February 2014 to March 2015, 85 subjects undergoing TAVI at 13 centres in Europe and Israelwere randomized to TriGuard protection vs. no protection. Subjects underwent neurologic and cognitive evaluation at baseline, pre-discharge and 30 days; cerebral diffusion-weighted magnetic resonance imaging was performed at 4 +/- 2 days post-procedure and at 30 days. Technical success, which included complete 3-vessel cerebral coverage, was achieved in 88.9% (40/45) of cases. The primary in-hospital procedural safety endpoint (death, stroke, life-threatening or disabling bleeding, stage2 or 3 acute kidney injury, or major vascular complications) occurred in 21.7% ofTriGuard and 30.8% of control subjects (P = 0.34). In the Per Treatment population (subjects with complete three-vessel cerebral coverage), TriGuard use was associated with greater freedom from new ischaemic brain lesions (26.9 vs. 11.5%), fewer new neurologic deficits detected by the National Institutes of Health Stroke Scale (3.1 vs. 15.4%), improved Montreal Cognitive Assessment (MoCA) scores, better performance on a delayed memory task (P = 0.028) at discharge, and a >2-fold increase in recovery of normal cognitive function (MoCA score >26) at 30 days. Conclusion TriGuard cerebral protection during TAVI is safe and complete cerebral vessel coveragewas achieved in 89% of subjects. In this exploratory study, subjects undergoing protected TAVI had more freedom from ischaemic brain lesions, fewer neurologic deficits, and improved cognitive function in some domains at discharge and 30 days compared with controls.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available