4.6 Article

Evidence that conflict regarding size of haemodynamic response to interventricular delay optimization of cardiac resynchronization therapy may arise from differences in how atrioventricular delay is kept constant

Journal

EUROPACE
Volume 17, Issue 12, Pages 1823-1833

Publisher

OXFORD UNIV PRESS
DOI: 10.1093/europace/euu374

Keywords

Heart failure; Cardiac resynchronization therapy; Physiology; Mechanisms

Funding

  1. British Heart Foundation [FS/11/92/29122, SP/10/002/28189, FS/08/027/24763, FS/13/44/30291, FS/10/038]
  2. British Heart Foundation [FS/08/027/24763, SP/10/002/28189, FS/10/38/28268, FS/13/44/30291, FS/11/92/29122, RG/10/11/28457, FS/14/27/30752] Funding Source: researchfish

Ask authors/readers for more resources

Aims Whether adjusting interventricular (VV) delay changes haemodynamic efficacy of cardiac resynchronization therapy (CRT) is controversial, with conflicting results. This study addresses whether the convention for keeping atrioventricular (AV) delay constant during VV optimization might explain these conflicts. Method and results Twenty-two patients in sinus rhythm with existing CRT underwent VV optimization using non-invasive systolic blood pressure. Interventricular optimization was performed with four methods for keeping the AV delay constant: (i) atrium and left ventricle delay kept constant, (ii) atrium and right ventricle delay kept constant, (iii) time to the first-activated ventricle kept constant, and (iv) time to the second-activated ventricle kept constant. In 11 patients this was performed with AV delay of 120 ms, and in 11 at AV optimum. At AV 120 ms, time to the first ventricular lead (left or right) was the overwhelming determinant of haemodynamics (13.75 mmHg at +/- 80 ms, P < 0.001) with no significant effect of time to second lead (0.47 mmHg, P = 0.50), P < 0.001 for difference. At AV optimum, time to first ventricular lead again had a larger effect (5.03 mmHg, P < 0.001) than timeto second (2.92 mmHg, P = 0.001), P = 0.02 for difference. Conclusion Time to first ventricular activation is the overwhelming determinant of circulatory function, regardless of whether this is the left or right ventricular lead. If this is kept constant, the effect of changing time to the second ventricle is small or nil, and is not beneficial. In practice, it may be advisable to leaveVVdelay at zero. Specifying howAV delay is kept fixed might make future VV delay research more enlightening.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available