3.8 Article

Assessment of dose uniformity around high dose rate 192Ir and 60Co stepping sources

Journal

RADIOLOGICAL PHYSICS AND TECHNOLOGY
Volume 10, Issue 4, Pages 454-463

Publisher

SPRINGER JAPAN KK
DOI: 10.1007/s12194-017-0418-1

Keywords

Brachytherapy; Stepping source; Dose uniformity

Funding

  1. Shahid Beheshti University of Medical Sciences

Ask authors/readers for more resources

This study aimed to evaluate dose uniformity for Ir-192 and Co-60 stepping sources. High dose rate Ir-192 and Co-60 stepping sources were simulated by the MCNPX Monte Carlo code. To investigate dose uniformity, treatment lengths of 30, 50, 100, and 150 mm with stepping distances of 3, 5, 7, and 10 mm were considered. Finally, dose uniformity for the Ir-192 and Co-60 stepping sources with increasing distances from the source were assessed at these treatment lengths and steps. The findings showed that the dose distribution was non-uniform for regions in close vicinity of the source, especially in the high source steps, but for most points at distances > 10 mm from the center of the source, the dose distribution was uniform. For most points, the dose uniformity increased with reduction of the source steps and increments of the transverse distance from the source. The dose non-uniformity was similar for most of the corresponding points of Co-60 and Ir-192 sources with the same treatment lengths and source steps, except at the distance of 150 mm. When using stepping technique for the treatment of tumors, more attention should be focused on treatment planning, especially with higher stepping distances and lower transverse distances from the source.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

3.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available