Mega Acer Kit® is more effective for warming the intravenous fluid than Ranger™ and ThermoSens® at 440 ml/h of infusion rate: an experimental performance study

Journal

KOREAN JOURNAL OF ANESTHESIOLOGY
Volume 70, Issue 4, Pages 456-461

Publisher

KOREAN SOC ANESTHESIOLOGISTS
DOI: 10.4097/kjae.2017.70.4.456

Keywords

Fluid temperature; Intravenous fluid warmer; Low flow rate; Mega Acer Kit; Ranger; ThermoSens

Categories

Funding

  1. Clinical Medicine Research Institute at Chosun University Hospital

Ask authors/readers for more resources

Background: Few studies have investigated the effectiveness of intravenous fluid warmers at low and moderate flow rates below 1,000 ml/h. In this study, we compared the effectiveness of three different fluid warmers at a low flow rate (440 ml/h). Methods: We experimentally investigated the fluid warming performances of Mega Acer Kit (R) (Group M, n = 10), Ranger (TM) (Group R, n = 10), and ThermoSens (R) (Group T, n = 10) at 440 ml/h for 60 min. All devices were set at a warming temperature of 41 degrees C with preheating for 10 min. Intravenous fluids were then delivered through them. The fluid temperature (primary endpoint) was measured at 76 cm from the device after infusion for 60 min. The expected decrease in mean body temperature (secondary endpoint) after 5 h infusion for a 70 kg patient (Delta MBT5) was also calculated. Results: The fluid temperature (mean [95% CI]) at 76 cm from the device, 60 minutes after the infusion was higher in group M (36.01[35.73-36.29]degrees C), compared to groups T (29.81 [29.38-30.24]degrees C) and R (29.12 [28.52-29.72]degrees C) (P < 0.001). The Delta MBT5 (mean [95% CI]) was significantly smaller in group M (-0.04 [-0.04 to -0.03]degrees C) than that in groups T (-0.27 [-0.28 to -0.29]degrees C; P < 0.001) and R (-0.30 [-0.32 to -0.27]degrees C; P < 0.001). However, none of the fluid warmers provided a constant normothermic temperature above 36.5 degrees C. Conclusions: Mega Acer Kit (R) was more effective in warming the intravenous fluid with the smallest expected change in the mean body temperature, compared to Ranger (TM) and ThermoSens (R), at a flow rate of 440 ml/h.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available