4.3 Article

Automated IMRT planning in Pinnacle

Journal

STRAHLENTHERAPIE UND ONKOLOGIE
Volume 193, Issue 12, Pages 1031-1038

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00066-017-1187-9

Keywords

Automated planning; Intensity-modulated radiotherapy; Head and neck cancer; Treatment planning; Auto-Planning

Funding

  1. Philips Healthcare, Fitchburg, WI, USA

Ask authors/readers for more resources

This study evaluates the performance and planning efficacy of the Auto-Planning (AP) module in the clinical version of Pinnacle 9.10 (Philips Radiation Oncology Systems, Fitchburg, WI, USA). Twenty automated intensity-modulated radiotherapy (IMRT) plans were compared with the original manually planned clinical IMRT plans from patients with oropharyngeal cancer. Auto-Planning with IMRT offers similar coverage of the planning target volume as the original manually planned clinical plans, as well as better sparing of the contralateral parotid gland, contralateral submandibular gland, larynx, mandible, and brainstem. The mean dose of the contralateral parotid gland and contralateral submandibular gland could be reduced by 2.5 Gy and 1.7 Gy on average. The number of monitor units was reduced with an average of 143.9 (18%). Hands-on planning time was reduced from 1.5-3 h to less than 1 h. The Auto-Planning module was able to produce clinically acceptable head and neck IMRT plans with consistent quality.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available