CologneAMS; Small gaseous CO2 samples; Radiocarbon; Gas ion source; Accelerator mass spectrometry
Categories
Funding
- Deutsches GeoForschungsZentrum GFZ
- Helmholtz-Zentrum Potsdam
Ask authors/readers for more resources
A second SO-110 B (Arnold et al., 2010) ion source was installed at the 6 MV CologneAMS for the measurement of gaseous samples. For the gas supply a dedicated device from Ionplus AG was connected to the ion source. Special effort was devoted to determine optimized operation parameters for the ion source, which give a high carbon current output and a high C-14(-) yield. The latter is essential in cases when only small samples are available. Additionally a modified immersion lens and modified target pieces were tested and the target position was optimized. (C) 2017 Elsevier B.V. All rights reserved.