Noise Rejection for Wearable ECGs Using Modeified Frequency Slice Wavelet Transform and Convolutional Neural Networks

Journal

IEEE ACCESS
Volume 7, Issue -, Pages 34060-34067

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/ACCESS.2019.2900719

Keywords

Wearable ECG; signal quality assessment (SQA); convolutional neural network (CNN); modified frequency slice wavelet transform (MFSWT)

Funding

  1. National Natural Science Foundation of China [81871444]
  2. Primary Research and Development Plan of Jiangsu Province [BE2017735]

Ask authors/readers for more resources

Progress in wearable techniques makes the long-term daily electrocardiogram (ECG) monitoring possible. However, the long-term wearable ECGs can be significantly contaminated by various noises, which affect the detection and diagnosis of cardiovascular diseases (CVDs). The situation becomes more serious for wearable ECG screening, where the data are huge, and doctors have no way to visually check the signal quality episode-by-episode. Therefore, automatic and accurate noise rejection for the wearable big-data ECGs is craving. This paper addressed this issue and proposed a noise rejection method for wearable ECGs based on the combination of modified frequency slice wavelet transform (MFSWT) and convolutional neural network (CNN). Wearable ECGs were recorded using the newly developed 12-lead Lenovo smart ECG vest with a sample rate of 500 Hz and a resolution of 16 bit. One thousand 10-s ECG segments were picked up and were manually labeled into three quality types: clinically useful segments with good signal quality (type A), clinically useful segments with poor signal quality (type B), and clinically useless segments (pure noises, type C). Each of the 1,000 10-s ECG segments were transformed into a 2-D time-frequency (T-F) image using the MFSWT, with a pixel size of 200x50. Then, the 2-D grayscale images from MFSWT were fed into a 13-layer CNN model for training the classification models. Results from the standard 5-folder cross-validation showed that the proposed combination method of MFSWT and CNN achieved a highest classification accuracy of 86.3%, which was higher than the comparable methods from continuous wavelet transform (CWT) and artificial neural networks (ANN). The combination of MFSWT and CNN also had a good calculation efficiency. This paper indicated that the combination of MFSWT and CNN is a potential method for automatic identification of noisy segments from wearable ECG recordings.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available