4.3 Article

Influence of operating parameters on the suture retention test for scaffolds in ophthalmology

Publisher

ELSEVIER SCIENCE BV
DOI: 10.1016/j.msec.2017.02.177

Keywords

Polycaprolactone; Cast; Biomaterials; Polymers; Tissue engineering; Suture; Suture retention; Characterization; Ophthalmology

Ask authors/readers for more resources

The strength of a suture used to attach a graft to a patient can be quantified using the suture retention test. In order to gain deeper insight on the influence of testing and application conditions testing parameters were varied. Different pull rates, suture and jaw positions, and suture materials were tested to show their influence on the result of the test. The results of the different states were analyzed using statistical tests. Based on the statistic analysis conditions for testing with a maximum of accuracy and as close as possible to in vivo conditions have been revealed. (C) 2017 Elsevier B.V. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available