3.9 Article

Full-field cavity enhanced microscopy techniques

Journal

JOURNAL OF PHYSICS-PHOTONICS
Volume 1, Issue 1, Pages -

Publisher

IOP Publishing Ltd
DOI: 10.1088/2515-7647/aae228

Keywords

multi-pass imaging; cavity-enhanced microscopy; contrast enhancement; optical imaging

Funding

  1. Gordon and Betty Moore Foundation
  2. Human Frontier Science Program
  3. National Research Foundation, Prime Minister'sOffice, Singapore
  4. Ministry of Education, Singapore under the Research Centres of Excellence programme

Ask authors/readers for more resources

The number of probe particles that is detected on a single pixel of a micrograph is finite, either due to source (low power), detector (low dynamic range) or specimen damage constraints. The sensitivity of an otherwise perfect microscope is then limited by the statistical fluctuations in the number of detected particles. It is thus crucial to strive for the optimal signal-to-noise ratio per detected photon. Here we analytically and numerically compare three different contrast enhancing techniques that are all based on self-imaging cavities: CW cavity enhanced microscopy, cavity ring-down microscopy and multi-pass microscopy. We show that all three schemes can lead to sensitivities beyond those achievable with a single pass.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

3.9
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available