4.5 Article

Two different techniques of manufacturing TMJ replacements - A technical report

Journal

JOURNAL OF CRANIO-MAXILLOFACIAL SURGERY
Volume 45, Issue 9, Pages 1432-1437

Publisher

CHURCHILL LIVINGSTONE
DOI: 10.1016/j.jcms.2017.06.003

Keywords

Custom own made implants; TMJ; DMLS; CNC

Funding

  1. Ministry of High Science of Polish Republic [503/5-061-02/503-51-001]
  2. Medical University of Lodz [0117/DIA/2014/43]

Ask authors/readers for more resources

Introduction: Presently, during the surgical treatment of the patients in maxillofacial surgery, one can use various medical implants. Moreover custom made implants are being used. Replacements may be fitted to the structure and shape of the human skull owing to CAD/CAM (custom aided design/manufacture) called customized implants. This study was aimed to report for the first time clinical material from which custom implants, using two different techniques, were manufactured to reconstruct the temporomandibular joint (TMJ). Material and methods: In this study, eleven patients with an average age of 54 years were included. All of the patients underwent TMJ reconstruction using direct metal laser sintering (DMLS) or computer numerical control milling (CNC) techniques for implant manufacture. Four of the eleven patients had a malignancy diagnosis, and seven had a benign diagnosis. Patients complained of hypomobility of the TMJ, facial asymmetry, pain and swelling of the preauricular region. Treatment included 7 CNC milled implants and 4 implants in DMLS. Results: More metallic implant parts with a rough surface were associated with the DMLS technique. Post operational, uneventful healing was observed in all clinical cases during an average of 26.8 months of follow-up. Three months post-operation, facial nerve palsy, swallowing disturbances and pain were not observed. Infections, allergic reactions to materials and re-ankylosis were also not observed. Replacements received correct forms and functions owing to the CAM techniques. Post-operational maximal interincisal opening improved (p < 0.01) and was not significantly related to preoperational opening, age, sex, diagnosis or adjuvant radiotherapy. Conclusion: Considering both methods, the feature that differentiates the manufacture technique is the more subtractive surface finishing required for the DMLS implant than the CNC implant. Both techniques resulted the same clinical outcomes and can be used successfully in patients with neoplastic lesions and other TMJ disorders. Unfortunately, DMLS is more vulnerable to fracture. (C) 2017 European Association for Cranio-Maxillo-Facial Surgery. Published by Elsevier Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available