4.6 Article

A Single-Center Experience With a Titanium Modular Neck Total Hip Arthroplasty

Journal

JOURNAL OF ARTHROPLASTY
Volume 32, Issue 8, Pages 2450-2456

Publisher

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.arth.2017.03.025

Keywords

modular neck; dislocations; hip arthroplasty; femoral neck fracture; stem

Categories

Ask authors/readers for more resources

Background: Added modular junction has been associated with implant-related failures. We report our experience with a titanium-titanium modular neck-stem interface to assess complications, possible clinical factors influencing use of neck modularity, and whether modularity reduced the incidence of dislocation. Methods: A total of 809 total hip arthroplasties completed between 2005 and 2012 from a prospectively collected database were reviewed. The mean follow-up interval was 5.7 years (3.3-10.3 years). Forty-five percent were male (360 of 809), and 55% were female (449 of 809). All stems were uncemented PROFEMUR TL (titanium, flat-tapered, wedge) or PROFEMUR Z (titanium, rectangular, dual-tapered) with a titanium neck. Results: Increased modularity (anteverted/retroverted and anteverted/retroverted varus/valgus (anteverted/ retroverted + anteverted/retroverted varus/valgus) was used in 39.4% (135 of 343) of cases using the posterior approach compared with 6.8% (20 of 293) of anterior and 23.7% (41 of 173) of lateral approaches. Four males sustained neck fractures at a mean of 95.5 months (69.3-115.6 months) after primary surgery. Overall dislocation rate was 1.1% (9 of 809). The posterior approach had both the highest utilization of increased modularity and the highest dislocation rate (2.3%), of which the most were recurrent. The anterior (0.3%) and lateral (0%) approaches had lower dislocation rates with no recurrences. Conclusion: At a mean 5.7 years, our experience demonstrates a low neck fracture (0.5%) and a low dislocation rate (1.1%). Use of increased modularity may not improve dislocation risk for the posterior approach. Continued surveillance of this group will be necessary to determine long term survivorship of this modular titanium implant. (C) 2017 Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available