4.7 Article

Trace element determination in high-salinity water using an online ion imprinted helical tube demineralizer combined with ICP-MS

Journal

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
Volume 35, Issue 7, Pages 1306-1315

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/d0ja00060d

Keywords

-

Funding

  1. Institutional Instrument Development Program of the Institute of Ecology and Environment of the Chinese Academy of Sciences [2016yyxm03]
  2. Chinese Major Science and Technology Program for Water Pollution Control and Treatment [2014ZX07405001]
  3. Edanz Group in China

Ask authors/readers for more resources

Matrix interference is a common problem in the measurement of trace elements in waterviaICP-MS. To solve this problem, a novel demineralizer was developed to combine with ICP-MS for determining trace elements in high-salinity water samples, which were filled with a novel ion imprinted material. In the analysis of water samples containing 120 mg L(-1)of Ca2+, Mg2+, and Na+, the recoveries of As, Cd, Cu, Mn, Ni, Pb, Sb, and Tl were all maintained between 80% and 120% during the 10 hours of the continuous operation of the demineralizer, with the trueness and repeatability for all elements standing below 3% and 5%, respectively. After the usage of the demineralizer, the detection limits of the method for As, Cd, Cu, Mn, Ni, Pb, Sb, and Tl were reduced to less than one-tenth of its original values. In the analysis of real water samples, a standard substance containing trace elements in the coastal water (CASS-5) and 5 offshore seawater samples were analyzed. The trueness and repeatability for all elements were below 5% and the recoveries were between 85-104%, which all met the requirements of analytical chemistry.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available