Effect of exposure time and moving the curing light on the degree of conversion and Knoop microhardness of light-cured resin cements

Journal

DENTAL MATERIALS
Volume 36, Issue 11, Pages E340-E351

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.dental.2020.08.016

Keywords

Light exposure technique; Light-curing unit; Exposure time; Degree of conversion; Microhardness; Resin cement; Ceramic veneer

Funding

  1. CNPq (National Council of Scientific and Technological Development)
  2. FAPEMIG (Foundation for Research Support of the State of Minas Gerais)
  3. (CAPES) Coordination for the Improvement of Higher Education Personnel

Ask authors/readers for more resources

Objective. To evaluate the effect of exposure time and moving the light-curing unit (LCU) on the degree of conversion (DC) and Knoop microhardness (KH) of two resin cements that were light-cured through ceramic. Methods. Two resin cements: AllCem Veneer APS (FGM) and Variolink Esthetic LC (Ivoclar Vivadent) were placed into a 0.3 mm thick matrix in 6 locations representing the canine to canine. The resins were covered with 0.5 mm thick lithium disilicate glass -ceramic (IPS e.max CAD, Ivoclar Vivadent). A motorized device moved the LCUs over the ceramic when the LCU was on. Two single-peak LCUs: Elipar DeepCure-L (3M Oral Care) and Emitter C (Schuster), and one multi-peak: Bluephase G2 (Ivoclar Vivadent) were used with 3 different exposure protocols: a localized exposure centered over each tooth for 10 or 40 s; moving the tip across the 6 teeth for a total exposure time of 10 or 40 s; and moving the tip across the 6 teeth resins for a total exposure time of 60 or 240 s. After 24 h, the DC and KH were measured on the top surfaces and the data was analyzed using three-way ANOVA and Tukey's tests (a = 0.05). Results. Interposition of 0.5 mm of ceramic reduced the irradiance received by the resin by approximately 50%. The 40 s localized exposure over each tooth always produced significantly higher DC and KH values. Moving the LCUs with a total exposure time of 10 s resulted in the lowest DC and KH. There was no beneficial effect on the DC or KH when the multi-peak (violet -blue) LCU (Elipar DeepCure-L or Bluephase G2), but the lower light output from a small tip LCU reduced the DC and KH values (Emitter C). Significance. Moving the LCUs when photo-curing light-cured resin cements is not recommended. This study showed that a single-peak LCU could activate a resin cement that uses IvocerinTM as well as the multi-peak LCU. (C) 2020 The Academy of Dental Materials. Published by Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available