4.6 Article

Identification of Rotors during Human Atrial Fibrillation Using Contact Mapping and Phase Singularity Detection: Technical Considerations

Journal

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
Volume 64, Issue 2, Pages 310-318

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBME.2016.2554660

Keywords

Atrial fibrillation; cardiac phase; contact mapping; Hilbert transform; instantaneous phase; phase singularities; unipolar electrograms

Funding

  1. European Network for Translational Research in AF (FP7 collaborative Project EUTRAF) [261057]
  2. CTMM COHFAR
  3. Dutch Heart Foundation
  4. Deutsches Zentrum fur Herzkreislaufforschung

Ask authors/readers for more resources

Objective: To explore technical challenges of phase singularity (PS) mapping during atrial fibrillation (AF) using direct contact electrograms. Methods: AF mapping was performed in high-density epicardial recordings of human paroxysmal (PAF) or persistent (PersAF) (N = 20 pts) AF with an array of 16 x 16 electrodes placed on atrial epicardium. PS points were detected using subsets of electrodes forming rings of varying sizes. Results: PS detection using a 2 x 2 electrode ring identified 0.88 +/- 1.00 PS/s in PAF group and 3.91 +/- 2.51 per s in PersAF group (p < 0.001) in 2.4 x 2.4 cm mapping area. All detected PS had a short lifespan with the longest being 1100 ms (6.8 rotations). Exploration of the PS detection in a numerical model demonstrated that at least eight electrodes are required to avoid frequent false positive PS detection due to chance. Application of a detection grid consisting a double ring of electrodes (2 x 2 and 4 x 4 rings) decreased the number of false positive detections. The double ring was more resilient to electrode swapping (with just three instances of false positives versus 4380 false positives using 2x2 ring). Conclusions: The number of detected rotors critically depends upon the parameters of the detection algorithm, especially the number of electrodes used to detect PS. Based on our results, we recommend double ring comprised of 2 x 2 and 4 x 4 grid of electrodes for robust rotor detection. Significance: Great methodological care has to be taken before equating detected PS with rotating waves and using PS detection algorithms to guide catheter ablation of AF.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available