4.4 Article

Atrioventricular Synchronous Pacing Using a Leadless Ventricular Pacemaker Results From the MARVEL 2 Study

Journal

JACC-CLINICAL ELECTROPHYSIOLOGY
Volume 6, Issue 1, Pages 94-106

Publisher

ELSEVIER
DOI: 10.1016/j.jacep.2019.10.017

Keywords

atrioventricular synchrony; AV block; leadless pacemaker

Funding

  1. Medtronic, Inc.
  2. Abbott
  3. Biotronik
  4. Boston Scientific
  5. Medtronic

Ask authors/readers for more resources

OBJECTIVES This study reports on the performance of a leadless ventricular pacemaker with automated, enhanced accelerometer-based algorithms that provide atrioventricular (AV) synchronous pacing. BACKGROUND Despite many advantages, leadless pacemakers are currently only capable of single-chamber ventricular pacing. METHODS The prospective MARVEL 2 (Micra Atrial tRacking using a Ventricular accELerometer 2) study assessed the performance of an automated, enhanced accelerometer-based algorithm downloaded to the Micra leadless pacemaker for up to 5 h in patients with AV block. The primary efficacy objective was to demonstrate the superiority of the algorithm to provide AV synchronous (VDD) pacing versus VVI-50 pacing in patients with sinus rhythm and complete AV block. The primary safety objective was to demonstrate that the algorithm did not result in pauses or heart rates of >100 beats/min. RESULTS Overall, 75 patients from 12 centers were enrolled; an accelerometer-based algorithm was downloaded to their leadless pacemakers. Among the 40 patients with sinus rhythm and complete AV block included in the primary efficacy objective analysis, the proportion of patients with >= 70% AV synchrony at rest was significantly greater with VDD pacing than with VVI pacing (95% vs. 0%; p < 0.001). The mean percentage of AV synchrony increased from 26.8% (median: 26.9%) during WI pacing to 89.2% (median: 94.3%) during VDD pacing. There were no pauses or episodes of oversensing-induced tachycardia reported during VDD pacing in all 75 patients. CONCLUSIONS Accelerometer-based atrial sensing with an automated, enhanced algorithm significantly improved AV synchrony in patients with sinus rhythm and AV block who were implanted with a leadless ventricular pacemaker. (C) 2020 The Authors. Published by Elsevier on behalf of the American College of Cardiology Foundation.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available