4.0 Article

Concept description and accuracy evaluation of a moldable surgical targeting system

Journal

JOURNAL OF MEDICAL IMAGING
Volume 8, Issue 1, Pages -

Publisher

SPIE-SOC PHOTO-OPTICAL INSTRUMENTATION ENGINEERS
DOI: 10.1117/1.JMI.8.1.015003

Keywords

cochlear implantation; image-guided surgery; micro-stereotactic frame; minimally invasive surgery; surgical template

Funding

  1. Hannover Medical School
  2. German Research Foundation (DFG) [RA 2751/4-1, EXC 2177/1, 390895286]

Ask authors/readers for more resources

The study introduces a concept for customizable guidance instrument in surgical procedures, utilizing bone cement to fabricate individualized surgical templates. Experimental results demonstrate high positioning accuracy and drilling precision, providing an easy-to-use and sterile approach with promising accuracy for surgical applications such as neurosurgery.
Purpose: We explain our concept for customization of a guidance instrument, present a prototype, and describe a set of experiments to evaluate its positioning and drilling accuracy. Methods: Our concept is characterized by the use of bone cement, which enables fixation of a specific configuration for each individual surgical template. This well-established medical product was selected to ensure future intraoperative fabrication of the template under sterile conditions. For customization, a manually operated alignment device is proposed that temporary defines the planned trajectory until the bone cement is hardened. Experiments (n = 10) with half-skull phantoms were performed. Analysis of accuracy comprises targeting validations and experiments including drilling in bone substitutes. Results: The resulting mean positioning error was found to be 0.41 +/- 0.30 mm at the level of the target point whereas drilling was possible with a mean accuracy of 0.35 +/- 0.30 mm. Conclusion: We proposed a cost-effective, easy-to-use approach for accurate instrument guidance that enables template fabrication under sterile conditions. The utilization of bone cement was proven to fulfill the demands of an easy, quick, and prospectively intraoperatively doable customization. We could demonstrate sufficient accuracy for many surgical applications, e.g., in neurosurgery. The system in this early development stage already outperforms conventional stereotactic frames and image-guided surgery systems in terms of targeting accuracy. (C) The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.0
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available