4.1 Article

Fluidic Considerations of Measuring Intracranial Pressure Using an Open External Ventricular Drain

Journal

CUREUS JOURNAL OF MEDICAL SCIENCE
Volume 13, Issue 5, Pages -

Publisher

SPRINGERNATURE
DOI: 10.7759/cureus.15324

Keywords

external ventricular drain; cerebrospinal fluid; intracranial hypertension; continuous measurement; traumatic brain injury; catheter; bioengineering; neuro-surgery; neuro-monitoring; neurology and critical care

Ask authors/readers for more resources

This study investigates the fluidic sources of error for ICP measurement with concurrent drainage in an EVD, showing that the error in pressure measurement increases linearly with flow rate but is not clinically significant. The predicted maximum error from fluidic effects within the drainage system is small compared to other sources of error in current ICP monitoring, allowing for clinical use.
Measurement of intracranial pressure (ICP) during cerebrospinal fluid (CSF) drainage with an external ventricular drain (EVD) typically requires stopping the flow during measurement. However, there may be benefits to simultaneous ICP measurement and CSF drainage. Several studies have evaluated whether accurate ICP measurements can be obtained while the EVD is open. They report differing outcomes when it comes to error, and hypothesize several sources of error. This study presents an investigation into the fluidic sources of error for ICP measurement with concurrent drainage in an EVD. Our experiments and analytical model both show that the error in pressure measurement increases linearly with flow rate and is not clinically significant, regardless of drip chamber height. At physiologically relevant flow rates (40 mL/hr) and ICP set points (13.6 - 31.3 cmH(2)O or 10 - 23 mmHg), our model predicts an underestimation of 0.767 cmH(2)O (0.56 mmHg) with no observed data point showing error greater than 1.09 cmH(2)O (0.8 mmHg) in our experiment. We extrapolate our model to predict a realistic worst-case clinical scenario where we expect to see a mean maximum error of 1.06 cmH(2)O (0.78 mmHg) arising from fluidic effects within the drainage system for the most resistive catheter. Compared to other sources of error in current ICP monitoring, error in pressure measurement due to drainage flow is small and does not prohibit clinical use. However, other effects such as ventricular collapse or catheter obstruction could affect ICP measurement under continuous drainage and are not investigated in this study.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available