3.8 Article

Ultra-low-dose non-contrast CT and CT angiography can be used interchangeably for assessing maximal abdominal aortic diameter

Journal

ACTA RADIOLOGICA OPEN
Volume 11, Issue 10, Pages -

Publisher

SAGE PUBLICATIONS LTD
DOI: 10.1177/20584601221132461

Keywords

Abdominal aortic aneurysm screening; Surveillance; Reproducibility; CT measurements; Agreement

Ask authors/readers for more resources

This study provides evidence that ultra-low-dose non-contrast CT (ULDNC-CT) exhibits similar accuracy and reproducibility to CT angiography (CTA) in assessing maximal abdominal aortic diameter. The findings suggest that ULDNC-CT can be used interchangeably with CTA for evaluating aortic sizes in the lower range.
Background: Routine CT scans may increasingly be used to document normal aortic size and to detect incidental abdominal aortic aneurysms. Purpose: To determine whether ultra-low-dose non-contrast CT (ULDNC-CT) can be used instead of the gold standard CT angiography (CTA) for assessment of maximal abdominal aortic diameter. Materials and Methods: This retrospective study included 50 patients who underwent CTA and a normal-dose non-contrast CT for suspected renal artery stenosis. ULDNC-CT datasets were generated from the normal-dose non-contrast CT datasets using a simulation technique. Using the centerline technique, radiology consultants (n = 4) and residents (n = 3) determined maximal abdominal aortic diameter. The limits of agreement with the mean (LOAM) was used to access observer agreement. LOAM represents how much a measurement by a single observer may plausibly deviate from the mean of all observers on the specific subject. Results: Observers completed 1400 measurements encompassing repeated CTA and ULDNC-CT measurements. The mean diameter was 24.0 and 25.0 mm for CTA and ULDNC-CT, respectively, yielding a significant but minor mean difference of 1.0 mm. The 95% LOAM reproducibility was similar for CTA and ULDNC-CT (2.3 vs 2.3 mm). In addition, the 95% LOAM and mean diameters were similar for CTA and ULDNC-CT when observers were grouped as consultants and residents. Conclusions: Ultra-low-dose non-contrast CT exhibited similar accuracy and reproducibility of measurements compared with CTA for assessing maximal abdominal aortic diameter supporting that ULDNC-CT can be used interchangeably with CTA in the lower range of aortic sizes.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

3.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available