4.4 Article

Calibration of the Latest Generation Superconducting Gravimeter iGrav-043 Using the Observatory Superconducting Gravimeter OSG-CT040 and the Comparisons of Their Characteristics at the Walferdange Underground Laboratory for Geodynamics, Luxembourg

Journal

PURE AND APPLIED GEOPHYSICS
Volume 180, Issue 2, Pages 629-641

Publisher

SPRINGER BASEL AG
DOI: 10.1007/s00024-021-02938-1

Keywords

Superconducting gravimeter; calibration factor; tides analyses; instrumental drift; gravity residuals

Ask authors/readers for more resources

This paper reports a study on the estimation of the calibration factor for a superconducting gravimeter. The results show that tidal analysis can be used to transfer the calibration factor between different gravimeters. Additionally, the study finds specific behavioral patterns in the instrumental drift of the iGrav-043.
In December 2019, the latest generation transportable superconducting gravimeter (SG) iGrav-043 purchased by the University of Bonn was installed in the Walferdange Underground Laboratory for Geodynamics (WULG) in the Grand Duchy of Luxembourg. In this paper, we estimate the calibration factor of the iGrav-043, which is essential for long-term gravity monitoring. We used simultaneously collected gravity data from the un-calibrated iGrav-043 and the calibrated Observatory superconducting gravimeter OSG-CT040 that operates continuously at WULG since 2002. The tidal analysis provides a simple way to transfer the calibration factor of one SG to the other. We then assess and compare tidal analyses, instrumental drifts and high frequency noises. After 20 years of continuous operation, the instrumental drift of the OSG-CT040 is almost zero. From 533 days of joint operation, we found that the instrumental drift of iGrav-043 exhibits a composite behavior: just after the setup and for two months a fast exponential decrease of 171 nm s(-2), then a linear with a rate of 66 nm s(-2) +/- 10 nm s(-2) per year. We suggest that a period of 3 months is sufficient for calibrating the iGrav. Accidental electrical power cuts triggered slight differences in the reaction and recovery of the OSG-CT040 and iGrav-043. However, it has been found that the long-term linear behavior of the drift was not affected.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available