AbsorbaTack™ vs. ProTack™ vs. sutures: a biomechanical analysis of cervical fixation methods for laparoscopic apical fixations in the porcine model

Journal

ARCHIVES OF GYNECOLOGY AND OBSTETRICS
Volume 307, Issue 3, Pages 863-871

Publisher

SPRINGER HEIDELBERG
DOI: 10.1007/s00404-022-06827-3

Keywords

Pelvic organ prolapse; Apical fixation; Cervicosacropexy; Polyvinylidene-fluoride; Uterosacral ligaments; Mesh fixation; Biomechanical analysis

Ask authors/readers for more resources

This study evaluated the biomechanical differences of using sutures, non-absorbable tacks, and absorbable tacks to fix polyvinylidene-fluoride tapes in an in vitro model. The results showed that the fixation with sutures can support a higher maximum load compared to non-absorbable tacks and absorbable tacks.
Purpose Treatment of pelvic organ prolapse (POP) often requires the use of synthetic mesh. In case of a novel and standardized bilateral apical fixation, both uterosacral ligaments are replaced by polyvinylidene-fluoride (PVDF) tapes. One of the main problems remains the fixation method, which should be stable, but also simple and quick to use. The current study evaluated biomechanical differences between the cervical tape fixation with sutures (group 1), non-absorbable tacks (group 2) and absorbable tacks (group 3) in an in vitro porcine model. Methods A total of 28 trials, conducted in three groups, were performed on porcine, fresh cadaver uteri. All trials were performed until mesh, tissue or fixation device failure occurred. Primary endpoints were the biomechanical properties maximum load (N), displacement at failure (mm) and stiffness (N/mm). The failure mode was a secondary endpoint. Results There was a significant difference between all three groups concerning the maximum load. Group 1 (sutures) supported a maximum load of 64 +/- 15 N, group 2 (non-absorbable tacks) yielded 41 +/- 10 N and group 3 (absorbable tacks) achieved 15 +/- 8 N. The most common failure mode was a mesh failure for group 1 and 2 and a fixation device failure for group 3. Conclusion The PVDF-tape fixation with sutures supports 1.5 times the load that is supported by non-absorbable tacks and 4.2 times the load that is supported by absorbable tacks. Nevertheless, there was also a stable fixation through tacks. Sutures are the significantly stronger and cheaper fixation device but may prolong the surgical time in contrast to the use of tacks.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available