Development and optimization of a frequency mixing sensor for adjacent samples quantitative detection on a lateral flow assay

Journal

BIOTECHNOLOGY JOURNAL
Volume -, Issue -, Pages -

Publisher

WILEY-V C H VERLAG GMBH
DOI: 10.1002/biot.202300190

Keywords

frequency-mixing technology (FMT); lateral flow assays (LFA); magnetic nanoparticle (MNP); multiplexing detection; point-of-care testing (POCT)

Ask authors/readers for more resources

This study optimized the sensor structure using mathematical physics modeling and finite element simulation to reduce signal interference and achieve multi-channel synchronous detection capability. The results showed that the optimized sensor can accurately identify adjacent samples with a smaller minimum spacing required.
Frequency-mixing technology has been widely used to precisely identify magnetic nanoparticles in applications of quantitative biomedical detection in recent years. Examples include immune adsorption, lateral flow assays (LFAs), and biomagnetic imaging. However, the signals of magnetic response generated by adjacent magnetic samples interfere with each other owing to the small spacing between them in applications involving multi-sample detection (such as the LFA and multiplexing detection). Such signal interference prevents the biosensor from obtaining characteristic peaks related to the concentration of adjacent biomarkers from the magnetic response signals. Mathematical and physical models of the structure of sensors based on frequency-mixing techniques were developed. The theoretical model was verified and its key parameters were optimized by using simulations. A new frequency-mixing magnetic sensor structure was then designed and developed based on the model, and the key technical problem of signal crosstalk between adjacent samples was structurally solved. Finally, standard cards with stable magnetic properties were used to evaluate the performance of the sensor, and strips of the gastrin-17 (G-17) LFA were used to evaluate its potential for use in clinical applications. The results show that the minimum spacing between samples required by the optimized sensor to accurately identify them was only about 4-5 mm, and the minimum detectable concentration of G-17 was 11 pg mL-1. This is a significant reduction in the required spacing between samples for multiplexing detection. The optimized sensor also has the potential for use in multi-channel synchronous signal acquisition, and can be used to detect synchronous magnetic signals in vivo. The frequency mixing technology (FMT) sensors have gained widespread application in biological quantitative detection and disease diagnosis. Nonetheless, the sensor's continuous detection performance is hindered due to signal interference between adjacent magnetic samples. In this study, mathematical physics modeling and finite element simulation were employed to optimize the sensor structure, aiming to reduce signal interference and achieve multi-channel synchronous detection capability.image

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available