The effect of patient's body weight, infusion connection point, and infusion pump position on intravenous multi-infusion drug delivery at low infusion rates suitable for premature neonates

Journal

JOURNAL OF PHARMACY AND PHARMACOLOGY
Volume -, Issue -, Pages -

Publisher

OXFORD UNIV PRESS
DOI: 10.1093/jpp/rgad108

Keywords

infusion pump; neonate; dosing error; drug delivery; intravenous infusion

Ask authors/readers for more resources

This study investigated the drug delivery rate through a multi-infusion system in premature neonates. The results showed that delayed drug delivery in this system may lead to dosing errors.
Introduction: Parenteral drug administration in the neonatal intensive care involves complex pharmacotherapy adjusted for the patient's weight, fluid allowance, and complex multi-infusion systems.Objectives: We investigated the delivery rate of a model drug through a multi-infusion system consisting of six intravenous infusions.Methods: Delivery rate of the model drug was determined after infusion initiation and termination. Measurements were collected spectrophotometrically in real time. Time to drug delivery and the amount of drug delivered were measured.Key findings: The longest time to drug delivery was observed for a 500 g neonate model with a distal infusion connection point and neutral pump position (337 +/- 30 min, P < 0.001). The shortest time was observed for a 1000 g neonate model in the combination of proximal infusion connection point and neutral pump position (18 +/- 12 min, P < 0.05). The expected 100% of the drug was delivered only in two combinations: 500 g and 1000 g neonate models, proximal infusion connection point and neutral pump position (100.4 +/- 4.7%, P = 0.819 and 100.2 +/- 2.7%, P = 0.874, respectively). While the least drug was delivered to a 500 g neonate model in the combination of distal infusion connection point and neutral pump position (27.5 +/- 5.8%, P < 0.001).Conclusions: Delayed drug delivery to premature neonates due to multi-infusion systems may compromise accurate drug administration and lead to dosing errors.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available