Point of care testing (POCT) of cholesterol in blood serum via a moving reaction boundary electrophoresis titration chip

Journal

ANALYTICAL METHODS
Volume 15, Issue 24, Pages 2971-2978

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/d3ay00550j

Keywords

-

Ask authors/readers for more resources

Researchers have developed a visual and portable method for cholesterol testing in blood samples based on an electrophoresis titration model. The method combines selective enzymatic reaction with electrophoretic titration to quantify cholesterol in blood serum. The experiments validate the feasibility of the method and demonstrate its high selectivity, portability, and visuality.
Cholesterol (CHO) in human blood is one of the most frequently and crucially quantified substances in diagnostic laboratories. However, visual and portable point of care testing (POCT) methods have been rarely developed for the bioassay of CHO in blood samples. Here, we developed an electrophoresis titration (ET) model, a chip device of similar to 60 grams, and a quantification method for the POCT of CHO in blood serum based on a moving reaction boundary (MRB). In this model, the selective enzymatic reaction is integrated with an ET chip for visual and portable quantification. At first, CHO reacted with cholesterol oxidase (CHOx) in the anode well, producing H2O2 and cholest-4-en-3-one in the solution. H2O2 further oxidized the colorless and chargeless leucocrystal violet (LCV) dye into violet colored positively charged crystal violet (CV+) and, under the influence of the electric field, the CV+ migrates in the ET channels and is titrated by the alkali of sodium hydroxide immobilized in the ET channels. The length covered by the MRB was measured as a function of the CHO content. The relevant experiments validated the feasibility of the model and method. Furthermore, the experiments revealed the high selectivity, portability, and visuality of the ET-MRB model, device, and method. Finally, the experiments showed a fair sensitivity of LOD of 5 mu M, good linearity of 10-1000 mu M (r(2) = 0.9919), fair stability (intra-day RSD of less than 5.09% and an inter-day RSD of less than 6.36%), and high recovery (99.4-105%). All the data and results indicate the potential of the ET-MRB model, chip device, and method for POCT of CHO in human blood samples.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available