4.4 Article

Establishment of Age- and Gender-Specific Reference Ranges for 36 Routine and 57 Cell Population Data Items in a New Automated Blood Cell Analyzer, Sysmex XN-2000

Journal

ANNALS OF LABORATORY MEDICINE
Volume 36, Issue 3, Pages 244-249

Publisher

KOREAN SOC LABORATORY MEDICINE
DOI: 10.3343/alm.2016.36.3.244

Keywords

Age; Automated blood cell analyzer; Gender; Item; Reference range; Sysmex XN-2000

Ask authors/readers for more resources

We established age- and gender-specific reference ranges for the 36 routine complete blood cell (CBC) and 57 cell population data (CPD) items in the Sysmex XN-2000 (Sysmex, Japan). In total, 280 peripheral blood samples were obtained from an equal number of healthy adults. Values for 36 routine items and 57 CPD items were obtained for each sample, and the results were categorized into six subgroups (N>39 in each subgroup) according to patient age (20-40, 41-60, and >60 yr) and gender (male and female), and compared with respect to age and gender differences. The majority of data items (22 of 36 routine CBC items and 44 of 57 CPD items) exhibited significant differences.(P <= 0.05) in their results with respect to age or gender, and several red cell-, lymphocyte-, and platelet-related data tended to decrease in women or older adults. These results provide a basis for establishing age- and gender-specific reference ranges for routine and CPD items in Sysmex XN-2000. Furthermore, these reference ranges could be used to determine clinical significance for new items of Sysmex XN-2000 in further studies.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available