4.2 Article

Correlation of Sysmex-XN9000 and CellaVision Advanced RBC software on anisocytosis

Journal

HEMATOLOGY
Volume 28, Issue 1, Pages -

Publisher

TAYLOR & FRANCIS LTD
DOI: 10.1080/16078454.2023.2248433

Keywords

Cellavision-DI60; blood automated analyzer; consistency analysis; advanced RBC software; anisocytosis

Categories

Ask authors/readers for more resources

This study evaluates the consistency of erythrocyte volume parameters between Sysmex-XN9000 and CellaVisionDI-60. The results show that both instruments have good accuracy and consistency in measuring heterogeneity degree. It is suggested to combine Sysmex-XN9000 and CellaVisionDI-60 for comprehensive anisocytosis evaluation in daily work.
Objective: This study aims to evaluate the consistency of heterogeneity degree of erythrocyte volume parameters between the blood automated analyzer Sysmex-XN9000 and the advanced red blood cell software CellaVisionDI-60.Method: 500 blood samples of volunteers were analyzed by Sysmex-XN9000 and CellaVision-DI60. The sensitivity, specificity, positive predictive value, negative predictive value, false positive rate, and false negative rate were evaluated. The consistency of all parameters was tested.Result: Taking the standard RBC group as the control group, the RBC parameters of the macrocytic and the microcytic group were compared. There was a statistical difference between the groups. ROC curve analysis showed that the best cutoff value of microcytic and of macrocytic affecting MCV were 4.1% and 5.7%, respectively. The best cutoff value of anisocytosis was 15.0%. The correlation coefficient between anisocytosis and red blood cell distribution width (RDW-CV) was 0.756. The sensitivity, specificity, positive predictive value and coincidence rate of anisocytosis were high. The false negative rate was 10.0%, and the false positive rate was 7.4%.Conclusion: All parameters of the degree of heterogeneity have good accuracy and consistency in the two instruments. Anisocytosis is with higher coincidence rate and positive predictive value. MIC and MAC have a good prediction on the increase or decrease of MCV. Although advanced RBC software's false negative and false positive rates are high, the red blood cell image system is more intuitive and time-saving in observing cells. Consequently, CellaVision-DI60 is suggested to combine with XN-9000 for judging the anisocytosis in daily work comprehensively.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available