Multicentric evaluation of eight glucose and four ketone blood meters

Journal

CLINICAL BIOCHEMISTRY
Volume 48, Issue 18, Pages 1310-1316

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2015.07.032

Keywords

Point-of-care; Glucose meter; Ketone meter; ISO 15197 norm; Diabetes; Glycemia

Ask authors/readers for more resources

Objectives: High precision meters for blood glycemia are mandatory for monitoring glucose status in patients, avoiding both hypo-and hyper-glycemia. Health care providers routinely used in both out-and inpatients point-of-care measurements of glucose and ketone. These measurements, frequently used for medical decisions, are known to be less accurate than those performed in laboratories. Our aim was to evaluate, within the frame of an Assistance Publique-Hopitaux de Paris (AP-HP) multicentric study, the performances of eight glucose and four ketone meters, either connected or non-connected to a laboratory software. Design and methods: Glucose meter accuracy, precision, correlation with plasma glucose determined in central laboratories and hematocrit interferences were determined according to the ISO 15197: 2003 norm. The same norm was applied for the determination of accuracy, precision and recovery of ketone meters for B-hydroxybutyrate measurements. Results and conclusion: Among those meters, seven were considered as acceptable for glucose measurement and two for ketone measurement. Since all meters do not fit clinically relevant criteria, meters' performances have to be evaluated before use in clinical practice. (C) 2015 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available