Galactose-1-phosphate uridyltransferase dried blood spot quality control materials for newborn screening tests

Journal

CLINICAL BIOCHEMISTRY
Volume 48, Issue 6, Pages 437-442

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2014.12.009

Keywords

Galactose-1-phosphate uridyltransferase; Galactosemia; Newborn screening; Blood spots; Quality control; Sample integrity

Funding

  1. Intramural CDC HHS [CC999999] Funding Source: Medline

Ask authors/readers for more resources

Objectives: We aimed to prepare dried-blood-spot (DBS) quality control (QC) materials for galactose-1-phosphate uridyltransferase (GALT), to evaluate their stability during storage and use, and to evaluate their performance in five DBS GALT test methods. Design and methods: We prepared and characterized GALT-normal and GALT-deficient DBS materials and compared GALT activities in DBSs after predetermined storage intervals at controlled temperatures and humidities. External evaluators documented the suitability of the DBS QC materials for use in five GALT test methods. Results: GALT activity losses from DBSs stored in low (<30%) humidity for 14 days at 45 degrees C, 35 days at 37 degrees C, 91 days at room temperature, 182 days at 4 degrees C, and 367 days at -20 degrees C were 54%, 53%, 52% 23%, and 7% respectively. In paired DBSs stored in high humidity (>50%) for identical intervals, losseswere: 45 degrees C-68%; 37 degrees C-79%; room temperature-72%, and 4 degrees C-63%. GALT activities in DBSs stored at 4 degrees C were stable throughout 19 excursions to room temperature. Twenty-five of 26 external evaluators, using five different GALT test methods, classified the GALT-deficient DBSs as outside normal limits. All evaluators classified the GALT-normal DBSs as within normal limits. Conclusions: Most of the GALT activity loss from DBSs stored at elevated or room temperature was attributable to the effects of storage temperature. Most of the loss from DBSs stored at 4 degrees C was attributable to the effects of elevated humidity. Loss from DBSs stored at -20 degrees C was insignificant. The DBS materials were suitable for monitoring performance of all five GALT test methods. (C) 2014 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available