Quantitative 177Lu SPECT imaging using advanced correction algorithms in non-reference geometry

Journal

PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS
Volume 32, Issue 12, Pages 1745-1752

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2016.09.014

Keywords

Quantitative imaging; Correction algorithms; Transmission-dependent convolution-subtraction correction algorithm; PRRT with Lu-177-labelled somatostatin analogue

Funding

  1. EURAMET MetroMRT project (European Commission, Belgium Seventh Framework Programme, Belgium) [217257]

Ask authors/readers for more resources

Peptide receptor therapy with Lu-177-labelled somatostatin analogues is a promising tool in the management of patients with inoperable or metastasized neuroendocrine tumours. The aim of this work was to perform accurate activity quantification of Lu-177 in complex anthropomorphic geometry using advanced correction algorithms. Acquisitions were performed on the higher Lu-177 photopeak (208 keV) using a Philips IRIX gamma camera provided with medium-energy collimators. System calibration was performed using a 16 mL Jaszczak sphere surrounded by non-radioactive water. Attenuation correction was performed using l-maps derived from CT data, while scatter and septal penetration corrections were performed using the transmission-dependent convolution-subtraction method. SPECT acquisitions were finally corrected for dead time and partial volume effects. Image analysis was performed using the commercial QSPECT software. The quantitative SPECT approach was validated on an anthropomorphic phantom provided with a home-made insert simulating a hepatic lesion. Quantitative accuracy was studied using three tumourto- background activity concentration ratios (6: 1, 9: 1, 14: 1). For all acquisitions, the recovered total activity was within 12% of the calibrated activity both in the background region and in the tumour. Using a 6: 1 tumour-to-background ratio the recovered total activity was within 2% in the tumour and within 5% in the background. Partial volume effects, if not properly accounted for, can lead to significant activity underestimations in clinical conditions. In conclusion, accurate activity quantification of Lu-177 can be obtained if activity measurements are performed with equipment traceable to primary standards, advanced correction algorithms are used and acquisitions are performed at the 208 keV photopeak using medium-energy collimators. (C) 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available