4.7 Article

Individually-addressable composite microneedle electrode array by mold-and-place method for glucose detection

Journal

SENSORS AND ACTUATORS B-CHEMICAL
Volume 401, Issue -, Pages -

Publisher

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2023.134884

Keywords

Microneedle; Electrochemical biosensor; ISF; Glucose; SWCNT

Ask authors/readers for more resources

This study presents a method to fabricate independently functioning microneedle (MN) electrodes with narrow intervals for high precision electrochemical sensing. The optimized mixture of photocurable polymer and single-wall carbon nanotubes was used to mold single composite MNs, which were then attached to pre-patterned electrodes. Plasma etching and electropolymerization were performed to enhance the electrochemical activity, and Prussian blue and glucose oxidase were electrodeposited on the MNs for glucose detection. The MN electrodes showed good sensitivity and linearity, and the feasibility of glucose detection was demonstrated in an in vivo mouse study.
Microneedle (MN)-based electrochemical biosensors enable minimally invasive and in situ monitoring of biomolecules from interstitial fluid (ISF) that is strongly correlated to their concentrations in the blood. When an array of MNs is used as either working or counter electrodes for electrochemical sensing, it is often difficult to have narrow spacing between them and this can lead to poor measurement accuracy due to increased resistance, low sensitivity, and slow response time. This study aims to develop a method to fabricate independently functioning MN electrodes with narrow intervals between them for high precision electrochemical sensing. A mixture of photocurable polymer (SU-8) and single-wall carbon nanotubes (SWCNTs) was optimized for pressure-assisted transfer (PAT) molding and electrical conductivity. Single composite MNs were molded by PAT molding, and then attached on pre-patterned electrodes. After the 'mold-and-place' of the composite MN structures, plasma etching and electropolymerization of PEDOT:PSS were performed to enhance the electrochemical activity. Prussian blue (PB) and glucose oxidase (GOx) were electrodeposited on surface-treated MNs to enable glucose detection. MN electrodes showed limit of detection (LOD) at 0.225 mM and linearity up to 20 mM. Finally, MN electrodes were constructed on a flexible polyimide film and demonstrated the feasibility of detecting glucose in an in vivo mouse study.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available