Numerical investigation reveals challenges in measuring the contrast recovery coefficients in PET

Journal

PHYSICS IN MEDICINE AND BIOLOGY
Volume 68, Issue 21, Pages -

Publisher

IOP Publishing Ltd
DOI: 10.1088/1361-6560/ad00fa

Keywords

positron emission tomography (PET); contrast recovery coefficient; NEMA NU 2; performance evaluation; image quality assessment

Ask authors/readers for more resources

This study quantifies the systematic uncertainties of CRC measurements in PET imaging and highlights the importance of scanner-independent parameters on the reliability and reproducibility of these measurements. The findings emphasize the trade-off between voxel size and noise-induced CRC fluctuations and suggest adjustments to the current NEMA IQ standards.
Objective. Contrast recovery coefficient (CRC) is essential for image quality (IQ) assessment in positron emission tomography (PET), typically measured according to the National Electrical Manufacturers Association (NEMA) NU 2 standard. This study quantifies systematic uncertainties of the CRC measurement by a numerical investigation of the effects from scanner-independent parameters like voxel size, region-of-interest (ROI) misplacement, and sphere position on the underlying image grid. Approach. CRC measurements with 2D and 3D ROIs were performed on computer-generated images of a NEMA IQ-like phantom, using voxel sizes of 1-4 mm for sphere diameters of 5-40 mm-first in absence of noise and blurring, then with simulated spatial resolution and image noise with varying noise levels. The systematic uncertainties of the CRC measurement were quantified from above variations of scanner-independent parameters. Subsampled experimental images of a NEMA IQ phantom were additionally used to investigate the impact of ROI misplacement at different noise levels. Main results. In absence of noise and blurring, systematic uncertainties were up to 28.8% and 31.0% with 2D and 3D ROIs, respectively, for the 10 mm sphere, with the highest impact from ROI misplacement. In all cases, smaller spheres showed higher uncertainties with larger voxels. Contrary to prior assumptions, the use of 3D ROIs did not exhibit less susceptibility for parameter changes. Experimental and computer-generated images both demonstrated considerable variations on individual CRC measurements when background coefficient-of-variation exceeded 20%, despite negligible effects on mean CRC. Significance. This study underscores the effect of scanner-independent parameters on reliability, reproducibility, and comparability of CRC measurements. Our findings highlight the trade-off between the benefits of smaller voxel sizes and noise-induced CRC fluctuations, which is not considered in the current version of the NEMA IQ standards. The results furthermore warrant adjustments to the standard to accommodate the advances in sensitivity and spatial resolution of current-generation PET scanners.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available