Impact of the confluence of cardiac motion and high spatial resolution on performance of ECG-gated imaging with an investigational photon-counting CT system: A phantom study

Journal

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2023.102683

Keywords

High -resolution; Photon -counting; ECG gated; Coronary; Phantom

Ask authors/readers for more resources

The study demonstrates that photon-counting CT can achieve higher spatial resolution, which helps improve cardiac imaging quality and reduce motion artifacts. High-resolution images perform better in vessel diameter and circularity compared to standard resolution, with slightly lower mutual information but still acceptable similarity.
Purpose: Photon-counting CT (PCCT) has higher spatial resolution that conventional EID CT which improves imaging of stationary coronary plaques and stents.. In this work, we evaluated the relationship between higher spatial resolution and motion acquisition on an investigational PCCT system.Methods: An investigational photon-counting CT scanner (Siemens CounT) with ECG gating was used to image a coronary tree phantom with models of healthy, stenotic, and stented arteries using a motion simulator. Images were acquired with matched clinical parameters at rest and 60 beats per minute. An additional set of high dose stationary images were averaged to generate a motion-free, reduced noise reference. Scans were completed at standard (0.5 mm2) and high-resolution (0.25 mm2). Motion images were reconstructed at multiple phases. Regions of interest were drawn around vessels and segmented. Percentage difference from the reference standard was evaluated for vessel diameter and circularity. Mutual information between the reference and stationary and motion datasets was used as a measure of volumetric similarity.Results: The stenotic vessel showed the most variation from the reference when compared to healthy or stented vessels. Compared to standard resolution, high-resolution images had lower bias for diameter (-0.012 +/- 0.19% vs-0.052 +/- 0.14%) and lower variability for circularity (-0.13 +/- 0.138% vs-0.12 +/- 0.144%). Both differences were found to be statistically significant. High-resolution images had a slightly lower mutual information (1.28) than standard resolution (1.31).Conclusion: The higher spatial resolution enabled by photon-counting CT can be harnessed for cardiac imaging as the benefits of high spatial resolution acquisitions remain relevant in the presence of motion.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available