The integrated relaxation pressure may not be an appropriate gold standard for deglutitive relaxation due to reliance on a single intragastric reference sensor

Journal

NEUROGASTROENTEROLOGY AND MOTILITY
Volume -, Issue -, Pages -

Publisher

WILEY
DOI: 10.1111/nmo.14635

Keywords

achalasia; esophagogastric junction; esophagus; manometry

Ask authors/readers for more resources

This study aimed to investigate the impact of selecting different gastric reference sensors on IRP measurement and diagnosis. The results showed that choosing different reference sensors could lead to variations in IRP values, resulting in inconsistent diagnoses in 28% of HRM studies. Selecting the correct gastric reference sensor is crucial for accurate test results and avoiding misdiagnosis.
BackgroundIntegrated relaxation pressure (IRP) calculation depends on the selection of a single gastric reference sensor. Variable gastric pressure readings due to sensor selection can lead to diagnostic uncertainty. This study aimed to examine the effect of gastric reference sensor selection on IRP measurement and diagnosis. MethodsWe identified high-resolution manometry (HRM) conducted between January and November 2017 with at least six intragastric reference sensors. IRP measurements and Chicago Classification 3.0 (CCv3) diagnoses were obtained for each of six gastric reference sensors. Studies were categorized as stable (no change in diagnosis) or variable (change in diagnosis with gastric reference selection). Variable diagnoses were further divided into variable normal/dysmotility (& GE;1 normal IRP measurement and & GE;1 CCv3 diagnosis), or variable dysmotility (& GE;1 CCv3 diagnosis, only elevated IRP measurements). Bland-Altman plots were used to compare IRP measurements within HRM studies. Key ResultsThe analysis included 100 HRM studies, among which 18% had variable normal/dysmotility, and 10% had variable dysmotility. The average IRP difference between reference sensors was 6.7 mmHg for variable normal/dysmotility and 5.9 mmHg for variable dysmotility. The average difference between the proximal-most and distal-most sensors was -1.52 mmHg (lower limit of agreement -10.03 mmHg, upper limit of agreement 7.00 mmHg). Conclusions & InferencesIRP values can vary greatly depending on the reference sensor used, leading to inconsistent diagnoses in 28% of HRM studies. Choosing the correct gastric reference sensor is crucial for accurate test results and avoiding misdiagnosis. Standardization of reference sensor selection or supportive testing for uncertain results should be considered.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available