4.5 Article

Coronary stent imaging in photon counting computed Tomography: Optimization of reconstruction kernels in a phantom

Journal

EUROPEAN JOURNAL OF RADIOLOGY
Volume 166, Issue -, Pages -

Publisher

ELSEVIER IRELAND LTD
DOI: 10.1016/j.ejrad.2023.110983

Keywords

Coronary computed tomography angiography; Photon counting computed tomography; Photon counting detector; Reconstruction kernel; Stent imaging; Cardiac stents

Ask authors/readers for more resources

This study investigates the effect of using optimized kernels on CT imaging of cardiac stents. The results show that in PCCT UHR mode, using optimized kernels significantly improves the assessment of the in-stent lumen of small cardiac stents. It is recommended to use optimized kernels for imaging of cardiac stents.
Purpose: Imaging stents and in-stent stenosis remains a challenge in coronary computed tomography angiography (CCTA). In comparison to conventional Computed Tomography, Photon Counting CT (PCCT) provides decisive clinical advantages, among other things by providing low dose ultra-high resolution imaging of coronary arteries. This work investigates the image quality in CCTA using clinically established kernels and those optimized for the imaging of cardiac stents in PCCT, both for in-vitro stent imaging in 400 & mu;m standard resolution mode (SRM) and 200 & mu;m Ultra High Resolution Mode (UHR).Methods: Based on experimental scans, vascular reconstruction kernels (Bv56, Bv64, Bv72) were optimized. In an established phantom, 10 different coronary stents with 3 mm diameter were scanned in the first clinically available PCCT. Scans were reconstructed with clinically established and optimized kernels. Four readers measured visible stent lumen, performed ROI-based density measurements and rated image quality.Results: Regarding the visible stent lumen, UHR is significantly superior to SRM (p < 0.001). In all levels, the optimized kernels are superior to the clinically established kernels (p < 0.001). One optimized kernel showed a significant reduction of noise compared to the clinically established kernels. Overall image quality is improved with optimized kernels.Conclusions: In a phantom study PCCT UHR with optimized kernels for stent imaging significantly improves the ability to assess the in-stent lumen of small cardiac stents. We recommend using UHR with an optimized sharp vascular reconstruction kernel (Bv72uo) for imaging of cardiac stent.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available