4.5 Article

Identifying patients at risk: multi-centre comparison of HeartMate 3 and HeartWare left ventricular assist devices

Journal

ESC HEART FAILURE
Volume 10, Issue 3, Pages 1656-1665

Publisher

WILEY PERIODICALS, INC
DOI: 10.1002/ehf2.14308

Keywords

Left ventricular assist device; LVAD; End-stage heart failure; Mechanical circulatory support; Centrifugal continuous flow pump

Ask authors/readers for more resources

This study compared the outcomes of patients using HeartWare (HVAD) and HeartMate 3 (HM3), and found that HM3 patients had better survival rates and lower incidence of heart failure. The results suggest that prophylactic exchange to HM3 is necessary for HVAD patients.
AimsSince the withdrawal of HeartWare (HVAD) from the global market, there is an ongoing discussion if and which patients require prophylactically exchange for a HeartMate 3 (HM3). Therefore, it is important to study outcome differences between HVAD and HM3 patients. Because centres differ in patient selection and standard of care, we performed a propensity score (PS)-based study including centres that implanted both devices and aimed to identify which HVAD patients are at highest risk. Methods and resultsWe performed an international multi-centre study (n = 1021) including centres that implanted HVAD and HM3. PS-matching was performed using clinical variables and the implanting centre. Survival and complications were compared. As a sensitivity analysis, PS-adjusted Cox regression was performed. Landmark analysis with conditional survival >2 years was conducted to evaluate long-term survival differences. To identify which HVAD patients may benefit from a HM3 upgrade, Cox regression using pre-operative variables and their interaction with device type was performed. Survival was significantly better for HM3 patients (P < 0.01) in 458 matched patients, with a median follow-up of 23 months. Within the matched cohort, HM3 patients had a median age of 58 years, and 83% were male, 80% of the HVAD patients were male, with a median age of 59 years. PS-adjusted Cox regression confirmed a significantly better survival for HM3 patients when compared with HVAD, with a HR of 1.46 (95% confidence interval 1.14-1.85, P < 0.01). Pump thrombosis (P < 0.01) and ischaemic stroke (P < 0.01) occurred less in HM3 patients. No difference was found for haemorrhagic stroke, right heart failure, driveline infection, and major bleeding. Landmark-analysis confirmed a significant difference in conditional survival >2 years after implantation (P = 0.03). None of the pre-operative variable interactions in the Cox regression were significant. ConclusionsHM3 patients have a significantly better survival and a lower incidence of ischaemic strokes and pump thrombosis than HVAD patients. This survival difference persisted after 2 years of implantation. Additional research using post-operative variables is warranted to identify which HVAD patients need an upgrade to HM3 or expedited transplantation.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available