This study assessed the performance of three immunoassay platforms (Roche Elecsys II, Abbott Alinity, and Siemens Centaur) compared to a liquid chromatography tandem mass spectrometry (LC-MS/MS) method in the 1 mg overnight dexamethasone suppression test (ONDST). The results showed that Roche and Siemens were more closely aligned with LC-MS/MS, while Abbott may decrease the sensitivity of ONDST.
Background The 1 mg overnight dexamethasone suppression test (ONDST) is recommended for the differential diagnosis of Cushing's syndrome and the investigation of adrenal incidentalomas. Despite documented variation in serum cortisol immunoassay performance, little has been published regarding its effect on the ONDST. Aims Assess the performance of three immunoassay platforms (Roche Elecsys II, Abbott Alinity & Siemens Centaur) when compared to a liquid chromatography tandem mass spectrometry (LC-MS/MS) method. Methods Samples (n = 77) sent to the laboratory as part of an ONDST were retrieved prior to disposal, anonymized, and analysed on all platforms. Samples with factors impacting immunoassay analysis quality were excluded. Results were statistically compared to an LC-MS/MS method that previously demonstrated excellent comparability to a candidate reference method. Results The Roche gen II showed a mean bias of -2.4 nmol/L and a Passing-Bablok fit of y = -0.9 + 0.97x. This was not affected by sex. The Abbott showed a mean bias -18.8 nmol/L, and a fit of y = -11.3 + 0.88x. This bias was -20.7 nmol/L in females versus -17.2 nmol/L in males. The Siemens had a mean bias of 2.3 nmol/L and a fit of y = 1.4 + 1.07x. This bias was 5.7 nmol/L in males versus -1.0 nmol/L in females. Conclusions Clinicians should be aware of the method-dependent variation that exists within serum cortisol analysis during the ONDSTs. Roche and Siemens aligned more closely with LC-MS/MS while the Abbot may cause a reduction in ONDST sensitivity. This data supports assay-specific cut-offs for the ONDST.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available