Computational Fluid Dynamics (CFD) and Finite Element Analysis (FEM) of a Customized Stent-Graft for Endovascular (EVAR) Treatment of Abdominal Aortic Aneurism (AAA)

Journal

APPLIED SCIENCES-BASEL
Volume 13, Issue 9, Pages -

Publisher

MDPI
DOI: 10.3390/app13095712

Keywords

stent-graft; patient specific; Computational Fluid Dynamics; finite element analysis

Ask authors/readers for more resources

In this study, a fluid-structure interaction analysis was conducted on an aneurysmatic aorta. The results showed the benefits of personalized stents/grafts in terms of blood flow and wall stress, supporting the shift towards customized stents/grafts in vascular surgery.
Background: The treatment of abdominal aortic aneurysm (AAA) is today commonly treated by inserting a stent-graft by the endovascular route, without resorting to open surgery. However, some clinical cases do not allow this less invasive approach, meaning that the stent-graft cannot be inserted and open surgery is used. Methods: In the study, we propose a fluid-structure interaction (FSI) analysis of an aneurysmatic aorta that could not be treated with Endovascular Aneurysm Repair (EVAR). The vessel is reconstructed through segmentation from CT scans and subsequently modeled on CAD software to create the surface and thickness of the vessel itself. Subsequently, we proceeded to carry out Computational Fluid Dynamics (CFD) and FSI simulation. We propose a computational study on a vessel geometry that is faithful to reality and customized. Results: Hemodynamic variable results of the carried out simulations indicate that low velocity and consequently very low WSS areas located in aneurysmal site are no longer found when conventional or patient-specific grafts are inserted. The wall stress distribution of aorta FEM analysis enabled the identification of the area at risk of failure, that is, in the posterior part of the aneurysm (similar to 10(7) Pa), while FSI analysis of the patient-specific graft led to a uniform von Mises stresses distribution (similar to 10(5) Pa), except for the junctions where peak stress occurred. Conclusion: The importance of this study is to highlight the benefits of the personalized stent/graft. As the authors expected, the study shows the numerous benefits of the customized stent/graft in terms of blood flow trend and wall stress compared to a traditional stent/graft by supporting the tendency to want to shift the target towards customized stents/grafts, also in the vascular surgery sector.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available