4.8 Article

Safety and performance of the second-generation drug-eluting absorbable metal scaffold in patients with de-novo coronary artery lesions (BIOSOLVE-II): 6 month results of a prospective, multicentre, non-randomised, first-in-man trial

Journal

LANCET
Volume 387, Issue 10013, Pages 31-39

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/S0140-6736(15)00447-X

Keywords

-

Funding

  1. Biotronik AG

Ask authors/readers for more resources

Background Absorbable scaffolds were designed to overcome the limitations of conventional, non-absorbable metal-based drug-eluting stents. So far, only polymeric absorbable scaffolds are commercially available. We aimed to assess the safety and performance of a novel second-generation drug-eluting absorbable metal scaffold (DREAMS 2G) in patients with de-novo coronary artery lesions. Methods We did this prospective, multicentre, non-randomised, first-in-man trial at 13 percutaneous coronary intervention centres in Belgium, Brazil, Denmark, Germany, Singapore, Spain, Switzerland, and the Netherlands. Eligible patients had stable or unstable angina or documented silent ischaemia, and a maximum of two de-novo lesions with a reference vessel diameter between 2.2 mm and 3.7 mm. Clinical follow-up was scheduled at months 1, 6, 12, 24, and 36. Patients were scheduled for angiographic follow-up at 6 months, and a subgroup of patients was scheduled for intravascular ultrasound, optical coherence tomography, and vasomotion assessment. All patients were recommended to take dual antiplatelet treatment for at least 6 months. The primary endpoint was in-segment late lumen loss at 6 months. We did analysis by intention to treat. This trial is registered with ClinicalTrials.gov, number NCT01960504. Findings Between Oct 8, 2013, and May 22, 2015, we enrolled 123 patients with 123 coronary target lesions. At 6 months, mean in-segment late lumen loss was 0.27 mm (SD 0.37), and angiographically discernable vasomotion was documented in 20 (80%) of 25 patients. Intravascular ultrasound assessments showed a preservation of the scaffold area (mean 6.24 mm(2) [SD 1.15] post-procedure vs 6.21 mm(2) [1.22] at 6 months) with a low mean neointimal area (0.08 mm(2) [0.09]), and optical coherence tomography did not detect any intraluminal mass. Target lesion failure occurred in four (3%) patients: one (<1%) patient died from cardiac death, one (<1%) patient had periprocedural myocardial infarction, and two (2%) patients needed clinically driven target lesion revascularisation. No definite or probable scaff old thrombosis was observed. Interpretation Our findings show that implantation of the DREAMS 2G device in de-novo coronary lesions is feasible, with favourable safety and performance outcomes at 6 months. This novel absorbable metal scaff old could be an alternative to absorbable polymeric scaffolds for treatment of obstructive coronary disease.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available