4.7 Article

Clinical evaluation of DIAGNOVIR SARS-CoV-2 ultra-rapid antigen test performance compared to PCR-based testing

Journal

SCIENTIFIC REPORTS
Volume 13, Issue 1, Pages -

Publisher

NATURE PORTFOLIO
DOI: 10.1038/s41598-023-31177-8

Keywords

-

Ask authors/readers for more resources

DIAGNOVIR is a rapid antigen testing system based on immune-fluorescence analysis, which can provide results within a minute. It has a sensitivity and specificity of 94% and 99.2% respectively, with 100% sensitivity among asymptomatic patients and the ability to detect SARS-CoV-2 within 5 days of symptom onset. Furthermore, DIAGNOVIR is not significantly affected by different variants of SARS-CoV-2 and has higher sensitivity and specificity, lower testing costs, and reduced turnaround time.
Coronavirus Disease-19 (COVID-19) is a highly contagious infectious disease caused by the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). The development of rapid antigen tests has contributed to easing the burden on healthcare and lifting restrictions by detecting infected individuals to help prevent further transmission of the virus. We developed a state-of-art rapid antigen testing system, named DIAGNOVIR, based on immune-fluorescence analysis, which can process and give the results in a minute. In our study, we assessed the performance of the DIAGNOVIR and compared the results with those of the qRT-PCR test. Our results demonstrated that the sensitivity and specificity of the DIAGNOVIR were 94% and 99.2%, respectively, with a 100% sensitivity and 96.97% specificity, among asymptomatic patients. In addition, DIAGNOVIR can detect SARS-CoV-2 with 100% sensitivity up to 5 days after symptom onset. We observed that the DIAGNOVIR Rapid Antigen Test's limit of detection (LoD) was not significantly affected by the SARS-CoV-2 variants including Wuhan, alpha (B1.1.7), beta (B.1.351), delta (B.1.617.2) and omicron (B.1.1.529) variants, and LoD was calculated as 8 x 10(2), 6.81 x 10(1.5), 3.2 x 10(1.5), 1 x 10(3), and 1 x 10(3.5) TCID50/mL, respectively. Our results indicated that DIAGNOVIR can detect all SARS-CoV-2 variants in just seconds with higher sensitivity and specificity lower testing costs and decreased turnover time.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available