4.6 Article

Head-Mounted Display for Clinical Evaluation of Neck Movement Validation with Meta Quest 2

Journal

SENSORS
Volume 23, Issue 6, Pages -

Publisher

MDPI
DOI: 10.3390/s23063077

Keywords

virtual reality; Meta Quest 2; neck; rehabilitation; validation

Ask authors/readers for more resources

This study validates the Meta Quest 2 HMD system as an alternative for screening neck movement in healthy people. The results demonstrate that the angles provided by the HMD system are valid to calculate the rotational angles of the neck in each of the three axes. This is significant for screening neck disorders in healthy individuals.
Neck disorders have a significant impact on people because of their high incidence. The head-mounted display (HMD) systems, such as Meta Quest 2, grant access to immersive virtual reality (iRV) experiences. This study aims to validate the Meta Quest 2 HMD system as an alternative for screening neck movement in healthy people. The device provides data about the position and orientation of the head and, thus, the neck mobility around the three anatomical axes. The authors develop a VR application that solicits participants to perform six neck movements (rotation, flexion, and lateralization on both sides), which allows the collection of corresponding angles. An InertiaCube3 inertial measurement unit (IMU) is also attached to the HMD to compare the criterion to a standard. The mean absolute error (MAE), the percentage of error (%MAE), and the criterion validity and agreement are calculated. The study shows that the average absolute errors do not exceed 1 degrees (average = 0.48 +/- 0.09 degrees). The rotational movement's average %MAE is 1.61 +/- 0.82%. The head orientations obtain a correlation between 0.70 and 0.96. The Bland-Altman study reveals good agreement between the HMD and IMU systems. Overall, the study shows that the angles provided by the Meta Quest 2 HMD system are valid to calculate the rotational angles of the neck in each of the three axes. The obtained results demonstrate an acceptable error percentage and a very minimal absolute error when measuring the degrees of neck rotation; therefore, the sensor can be used for screening neck disorders in healthy people.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available