A novel biomimetic artificial intervertebral disc (bioAID) replacement implant has been developed and evaluated for its design safety based on biomechanical parameters. The results indicate that the implant can withstand shear-compression loads and mimic the behavior of natural disc under a diurnal loading pattern.
A novel biomimetic artificial intervertebral disc (bioAID) replacement implant has been developed containing a swelling hydrogel representing the nucleus pulposus, a tensile strong fiber jacket as annulus fibrosus and tita-nium endplates with pins to primarily secure the device between the vertebral bodies. In this study, the design safety of this novel implant was evaluated based on several biomechanical parameters, namely compressive strength, shear-compressive strength, risk of subsidence and device expulsion as well as identifying the diurnal creep-recovery characteristics of the device. The bioAID remained intact up to 1 kN under static axial compression and only 0.4 mm of translation was observed under a compressive shear load of 20 N. No subsidence was observed after 0.5 million cycles of si-nusoidal compressive loading between 50 and 225 N. After applying 400 N in antero-posterior direction under 100 N axial compressive preload, approximately 2 mm displacement was found, being within the range of displacements reported for other commercially available cervical disc replacement devices. The diurnal creep recovery behavior of the bioAID closely resembled what has been reported for natural intervertebral discs in literature. Overall, these results indicate that the current design can withstand (shear-compression loads and is able to remain fixed in a mechanical design resembling the vertebral bodies. Moreover, it is one of the first implants that can closely mimic the poroelastic and viscoelastic behavior of natural disc under a diurnal loading pattern.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available