Exploiting SMART pass filter deactivation detection to minimize inappropriate subcutaneous implantable cardioverter defibrillator therapies: a real-world single-centre experience and management guide

Journal

EUROPACE
Volume -, Issue -, Pages -

Publisher

OXFORD UNIV PRESS
DOI: 10.1093/europace/euad040

Keywords

Subcutaneous ICD; S-ICD; Oversensing; TWOS; Defibrillation; Programming; Inappropriate shocks

Ask authors/readers for more resources

The aim of this study was to evaluate the impact of the SMART Pass (TM) algorithm on inappropriate therapy (IT) in subcutaneous internal cardiac defibrillator (S-ICD) and to explore the association between SP deactivation and IT. The results showed that SP deactivation was significantly associated with an increased risk of IT, and changing the sensing vector was successful in preventing IT.
Aims The SMART Pass (TM) (SP) algorithm is a high-pass filter that aims to reduce inappropriate therapy (IT) in subcutaneous internal cardiac defibrillator (S-ICD), but SP can deactivate due to low amplitude sensed R waves or asystole. The association between IT and SP deactivation and management strategies were evaluated, hypothesizing SP deactivation increases the risk of IT and device re-programming, or lead/generator re-positioning could reduce this risk. Methods and results Retrospective single-centre audit of Emblem (TM) S-ICD devices implanted 2016 to 2020 utilizing health records and remote monitoring data. Cox regression models evaluated associations between SP deactivation and IT. A total of 348 patients (27 +/- 16.6 months follow-up) were studied: 73% primary prevention. Thirty-eight patients (11.8%) received 83 shocks with 27 patients (7.8%) receiving a total of 44 IT. Causes of IT were oversensing (98%) and aberrantly conducted atrial fibrillation (2%). SP deactivation occurred in 32 of 348 patients (9%) and was significantly associated with increased risk of IT (hazard ratio 5.36, 95% CI 2.37-12.13). SP deactivation was due to low amplitude R waves (94%), associated with a higher defibrillation threshold at implant and presence of arrhythmogenic right ventricular cardiomyopathy. No further IT occurred 16 +/- 15.5 months after corrective interventions, with changing the sensing vector being successful in 59% of cases. Conclusion To reduce the risk of IT, the cause of the SP deactivation should be investigated, and appropriate reprogramming, device, or lead modifications made. Utilizing the alert for SP deactivation and electrograms could pro-actively prevent IT.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available