4.6 Article

Complete sequence verification of plasmid DNA using the Oxford Nanopore Technologies' MinION device

Journal

BMC BIOINFORMATICS
Volume 24, Issue 1, Pages -

Publisher

BMC
DOI: 10.1186/s12859-023-05226-y

Keywords

MinION; Cell therapy; Plasmid sequencing; Long-read sequencing

Ask authors/readers for more resources

This study presents a cost-effective and accurate plasmid sequencing method using the MinION device from Oxford Nanopore Technologies as an alternative to capillary-based sequencing. The procedure allows for the identification of plasmid identity and mutations, achieving high accuracy in consensus sequence generation. This pipeline offers significant cost savings compared to outsourcing clinical-grade sequencing, providing accessible high-quality plasmid sequence verification.
BackgroundSequence verification is essential for plasmids used as critical reagents or therapeutic products. Typically, high-quality plasmid sequence is achieved through capillary-based Sanger sequencing, requiring customized sets of primers for each plasmid. This process can become expensive, particularly for applications where the validated sequence needs to be produced within a regulated and quality-controlled environment for downstream clinical research applications.ResultsHere, we describe a cost-effective and accurate plasmid sequencing and consensus generation procedure using the Oxford Nanopore Technologies' MinION device as an alternative to capillary-based plasmid sequencing options. This procedure can verify the identity of a pure population of plasmid, either confirming it matches the known and expected sequence, or identifying mutations present in the plasmid if any exist. We use a full MinION flow cell per plasmid, maximizing available data and allowing for stringent quality filters. Pseudopairing reads for consensus base calling reduces read error rates from 5.3 to 0.53%, and our pileup consensus approach provides per-base counts and confidence scores, allowing for interpretation of the certainty of the resulting consensus sequences. For pure plasmid samples, we demonstrate 100% accuracy in the resulting consensus sequence, and the sensitivity to detect small mutations such as insertions, deletions, and single nucleotide variants. In test cases where the sequenced pool of plasmids contains subclonal templates, detection sensitivity is similar to that of traditional capillary sequencing.ConclusionsOur pipeline can provide significant cost savings compared to outsourcing clinical-grade sequencing of plasmids, making generation of high-quality plasmid sequence for clinical sequence verification more accessible. While other long-read-based methods offer higher-throughput and less cost, our pipeline produces complete and accurate sequence verification for cases where absolute sequence accuracy is required.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available