Assessment of Intra-Abdominal Pressure with a Novel Continuous Bladder Pressure Monitor-A Clinical Validation Study

Journal

LIFE-BASEL
Volume 13, Issue 2, Pages -

Publisher

MDPI
DOI: 10.3390/life13020384

Keywords

intra-abdominal hypertension; abdominal compartment syndrome; intra-abdominal pressure monitoring; foley manometer

Ask authors/readers for more resources

This study aimed to test the accuracy of a novel continuous intra-abdominal pressure (IAP) monitor. The results showed a positive correlation between the novel monitor and the gold-standard foley manometer. Good agreement was found between the two methods, with small mean bias and standard deviation. However, further studies are needed to validate the performance in a wider range of pressures.
Introduction: Intra-abdominal hypertension and the resulting abdominal compartment syndrome are serious complications of severely ill patients. Diagnosis requires an intra-abdominal pressure (IAP) measurement, which is currently cumbersome and underused. We aimed to test the accuracy of a novel continuous IAP monitor. Methods: Adults having laparoscopic surgery and requiring urinary catheter intra-operatively were recruited to this single-arm validation study. IAP measurements using the novel monitor and a gold-standard foley manometer were compared. After anesthesia induction, a pneumoperitoneum was induced through a laparoscopic insufflator, and five randomly pre-defined pressures (between 5 and 25 mmHg) were achieved and simultaneously measured via both methods in each participant. Measurements were compared using Bland-Altman analysis. Results: In total, 29 participants completed the study and provided 144 distinct pairs of pressure measurements that were analyzed. A positive correlation between the two methods was found (R-2 = 0.93). There was good agreement between the methods, with a mean bias (95% CI) of -0.4 (-0.6, -0.1) mmHg and a standard deviation of 1.3 mmHg, which was statistically significant but of no clinical importance. The limits of agreement (where 95% of the differences are expected to fall) were -2.9 and 2.2 mmHg. The proportional error was statistically insignificant (p = 0.85), suggesting a constant agreement between the methods across the range of values tested. The percentage error was 10.7%. Conclusions: Continuous IAP measurements using the novel monitor performed well in the clinical setup of controlled intra-abdominal hypertension across the evaluated range of pressures. Further studies should expand the range to more pathological values.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available