Temporal, biomechanical evaluation of a novel, transcatheter polymeric aortic valve in ovine aortic banding model

Journal

FRONTIERS IN CARDIOVASCULAR MEDICINE
Volume 9, Issue -, Pages -

Publisher

FRONTIERS MEDIA SA
DOI: 10.3389/fcvm.2022.977006

Keywords

aortic valve stenosis; transcatheter aortic valve replacement (TAVR); heart valve prosthesis; polymeric valve; preclinical in vivo study

Funding

  1. National Centre for Research and Development [STRATEGMED1/233166/6/NCBR/2014]

Ask authors/readers for more resources

This preclinical study evaluated the functionality, durability, and biocompatibility of a novel transcatheter heart valve. The results showed good hemodynamic performance and durability of the valve in a surgical model. Further studies with longer follow-ups are required.
ObjectivesThe aim of the study is to evaluate the functionality, durability, and temporal biocompatibility of a novel, balloon-expandable polymeric transcatheter heart valve (ATHV) system (InFlow, CardValve Consortium, Poland). Along with expanding TAVI indications, the demand for new transcatheter valves is increasing. MethodsA surgical ascending aortic banding model was created in 20 sheep. Two weeks later, 16 sheep were implanted with ATHV systems (15-16F). Three animals were euthanized after a 30-day follow-up, four animals after a 90-day follow-up, and six animals after a 180-day follow-up. A follow-up transthoracic echocardiography (TTE) was performed. ResultsThere was one procedure-related (6,25%) and two model-related deaths (12,5%; banding site calcification with subsequent infection originating externally from banding). TTE revealed the flow gradients (max/average) of 30,75/17,91; 32,57/19,21; and 21,34/10,63 mmHg at 30, 90, and 180 days, respectively. There were two cases of low-degree regurgitation after 180 days with no perivalvular leak observed. Histopathological analysis showed no valve degeneration at terminal follow-up with optimal healing. Small thrombi were present at the aortic wall adjacent to the base of the leaflets, and between the aortic wall and the stent in most of the valves; however, leaflets remained free from thrombi in all cases. Scanty calcifications of leaflets were reported in three animals evaluated 180 days after implantation. ConclusionThis preclinical study in the aortic banding model showed good hemodynamic performance, durability, and biocompatibility of the novel ATHV. Furthermore, regulatory studies with longer follow-ups are warranted.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available