4.3 Article

Efficacy and Safety of Transvenous Lead Extraction at the Time of Upgrade from Pacemakers to Cardioverter-Defibrillators and Cardiac Resynchronization Therapy

Publisher

MDPI
DOI: 10.3390/ijerph20010291

Keywords

device upgrade; transvenous lead extraction; prophylaxis of lead abandonment; restoration of venous access

Ask authors/readers for more resources

This study aimed to assess the complexity, risk, and outcome of transvenous lead extraction (TLE) among patients undergoing pacemaker upgrade to ICDs and CRTs. The results showed that TLE was the leading indication for upgrade in only 36-66% of patients, and the success rate of the upgrade procedure was 42.0-57.6%. Overall, the long-term survival rate was lower in the CRT-upgrade group compared to the non-upgrade group.
Background: Upgrading from pacemakers to ICDs and CRTs is a difficult procedure, and often, transvenous lead extraction (TLE) is necessary for venous access. TLE is considered riskier in patients with multiple diseases. We aimed to assess the complexity, risk, and outcome of TLE among CRT and ICD candidates. Methods: We analyzed clinical data from 2408 patients undergoing TLE between 2006 and 2021. There were 138 patients upgraded to CRT-D, 33 patients upgraded to CRT-P and 89 individuals upgraded to ICD versus 2148 patients undergoing TLE for other non-infectious indications. Results: The need for an upgrade was the leading indication for TLE in only 36-66% of patients. In 42.0-57.6% of patients, the upgrade procedure could be successfully done only after reestablishing access to the occluded vein. All leads were extracted in 68.1-76.4% of patients, functional leads were retained in 20.2-31.9%, non-functional leads were left in place in 0.0-1.1%, and non-functional superfluous leads were extracted in 3.6-8.4%. The long-term survival rate of patients in the CRT-upgrade group was lower (63.8%) than in the non-upgrade group (75.2%). Conclusions: Upgrading a patient from an existing pacemaker to an ICD/CRT is feasible in 100% of cases, provided that TLE is performed for venous access. Major complications of TLE at the time of device upgrade are rare and, if present do not result in death.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available