Next generation of dielectric barrier discharge hydride atomizers for atomic absorption spectrometry: A case study on Pb, Bi, Se and Te

Journal

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.sab.2022.106577

Keywords

Dielectric barrier discharge hydride atomizer; Selenium; Tellurium; Lead; Bismuth

Categories

Ask authors/readers for more resources

The performance of hydride atomizers based on dielectric barrier discharge (DBD) was assessed using atomic absorption spectrometry. Two designs of DBD atomizers with different electrode structures and power sources were compared. Nafion tubes were identified as effective gas phase dryers. Sensitivity varied among different analytes and atomizer configurations, with the best performance observed in a DBD atomizer with sputtered electrodes and square-wave modulation of high voltage.
The performance of hydride atomizers based on dielectric barrier discharge (DBD) was evaluated by atomic absorption spectrometry as a detector employing Pb, Bi, Se and Te as model analytes. Two designs of DBD atomizers either with glued or sputtered electrodes were investigated in combination with sinusoidal or square-wave modulated high voltage power supply sources. Moreover, the effect of gas phase dryers on analyte response was studied. Nafion tubes were found as universal dryers capable of efficient water vapor removal with no losses of analyte. Sensitivity was compared among individual DBD configurations being related also to conventional externally heated quartz tube (multi)atomizers (MM)QTA. For Te and Se comparable sensitivities were reached in all configurations of DBD atomizers and MMQTA. On the contrary, QTA provided three times higher sensitivity for Bi and Pb than DBD atomizers. Sensitivity reached among DBD atomizers was the same regardless of their electrode design and high voltage waveform modulation for Se, Te and Bi. Differences in sensitivity were observed for Pb depending on DBD atomizer configuration. Here the best sensitivity was found in DBD with sputtered electrodes coupled to power source with square-wave modulation of high voltage. It was twice higher than that reached in the DBD atomizer with glued electrodes powered by sine-wave voltage. The sensitivity in DBD atomizers correlates negatively with the amount of analyte deposited in the atomizer during atomization. >90% of Bi and Pb introduced into the system is deposited while the deposited fraction of Te and Se, respectively, decreases to 60% and 35%.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available