4.7 Article

Phantom-based evaluation of yttrium-90 datasets using biograph vision quadra

Journal

Publisher

SPRINGER
DOI: 10.1007/s00259-022-06074-3

Keywords

Whole-body PET; CT; Yttrium-90; NEMA; Image quality; Quadra

Ask authors/readers for more resources

This study evaluates the image quality characteristics of two NEMA phantoms with Y-90 on a long AFOV PET/CT and identifies the optimized reconstruction setup for imaging patients with hepatocellular carcinoma after Y-90 radioembolization. The results show that a 20-minute reconstruction time provides images with acceptable signal-to-noise ratio.
Purpose: The image quality characteristics of two NEMA phantoms with yttrium-90 (Y-90) were evaluated on a long axial field-of-view (AFOV) PET/CT. The purpose was to identify the optimized reconstruction setup for the imaging of patients with hepatocellular carcinoma after Y-90 radioembolization.Methods:Two NEMA phantoms were used, where one had a 1:10 sphere to background activity concentration ratio and the second had cold background. Reconstruction parameters used are as follows: iterations 2 to 8, Gaussian filter 2- to 6-mm full-width-at-half-maximum, reconstruction matrices 440 x 440 and 220 x 220, high sensitivity (HS), and ultra-high sensitivity (UHS) modes. 50-, 40-, 30-, 20-, 10-, and 5-min acquisitions were reconstructed. The measurements included recovery coefficients (RC), signal-to-noise ratio (SNR), background variability, and lung error which measures the residual error in the corrections. Patient data were reconstructed with 20-, 10-, 5-, and 1-min time frames and evaluated in terms of SNR.Results: The RC for the hot phantom was 0.36, 0.45, 0.53, 0.63, 0.68, and 0.84 for the spheres with diameters of 10, 13, 17, 22, 28, and 37 mm, respectively, for UHS 2 iterations, a 220 x 220 matrix, and 50-min acquisition. The RC values did not differ with acquisition times down to 20 min. The SNR was the highest for 2 iterations, measured 11.7, 16.6, 17.6, 19.4, 21.9, and 27.7 while the background variability was the lowest (27.59, 27.08, 27.36, 26.44, 30.11, and 33.51%). The lung error was 18%. For the patient dataset, the SNR was 19%, 20%, 24%, and 31% higher for 2 iterations compared to 4 iterations for 20-, 10-, 5-, and 1-min time frames, respectively.Conclusions: This study evaluates the NEMA image quality of a long AFOV PET/CT scanner with Y-90. It provides high RC for the smallest sphere compared to other standard AFOV scanners at shorter scan times. The maximum patient SNR was for 2 iterations, 20 min, while 5 min delivers images with acceptable SNR.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available