The safety and impact of the advanced hybrid closed-loop (AHCL) system on glycemic outcome in 2- to 6-year-old children with type 1 diabetes and the diabetes distress of caregivers were evaluated. The study found that after 12 weeks of using the AHCL system, glycemic control in children improved and diabetes distress in caregivers decreased.
Objective: The safety and impact of the advanced hybrid closed-loop (AHCL) system on glycemic outcome in 2- to 6-year-old children with type 1 diabetes and the diabetes distress of caregivers were evaluated.Research Design and Methods: This was an open-label prospective study (n = 35) with historical controls matched by treatment unit, diabetes duration, age, gender, and baseline treatment modality. The inclusion criteria were (1) type 1 diabetes diagnosis >6 months, (2) total daily dose of insulin >= 8 U/day, (3) HbA1c <10% (85 mmol/mol), and (4) capability to use insulin pump and continuous glucose monitoring. The MiniMed 780G (TM) AHCL in SmartGuard (TM) Mode was used for 12 weeks. Parental diabetes distress was evaluated with a validated Problem Areas In Diabetes-Parent, revised (PAID-PR) survey.Results: No events of diabetic ketoacidosis or severe hypoglycemia occurred. Between 0 and 12 weeks, HbA1c (mean change = -2.7 mmol/mol [standard deviation 5.7], P = 0.010), mean sensor glucose value (SG) (-0.8 mmol/L [1.0], P < 0.001), and time above range (TAR) (-8.6% [9.5], P < 0.001) decreased and time in range (TIR) (8.3% [9.3], P < 0.001) increased significantly, whereas no significant change in time below range (TBR) was observed. At the same time, PAID-PR score decreased from 37.5 (18.2) to 27.5 (14.8) (P = 0.006).Conclusions: MiniMed 780G (TM) AHCL is a safe system and 12-week use was associated with improvements in glycemic control in 2- to 6-year-old children with type 1 diabetes. In addition, AHCL is associated with a reduction in parental diabetes distress after 12-week use. ClinicalTrials.gov registration number: NCT04949022

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available